Používateľská príručka modulu METER BARO

METER BARO Module User Guide

Logo METERModul merača barometra

METER-BARO-Module -product

PRÍRUČKA INTEGRÁTORA BARO

POPIS SNÍMAČA

Modul BARO je presný barometer na kompenzáciu meraní matricového potenciálu tenziometrov TEROS 31 a TEROS 32. Modul BARO sa môže použiť ako samostatný senzor na kompenzáciu jedného alebo viacerých tenziometrov na mieste merania alebo ako digitálno-analógový prevodník na kompenzáciu hodnoty pripojeného TEROS 31 alebo TEROS 32 a prevod signálu SDI-12 na analógový objemový signál.tagvýstup e (iba 8-pinová verzia). Kombináciu modulu BARO a TEROS 32 je možné použiť ako náhradu za tenziometer T8. Podrobnejší popis spôsobu, akým tento senzor vykonáva merania, nájdete v používateľskej príručke k modulu BARO.

METER-BARO-Module- (1)

APLIKÁCIE

  • Barometric pressure measurement
  • Barometric compensation of matric potential measurements
  • Digital/analog converter for directly connected TEROS 31 and TEROS 32 tensiometers
  • Appropriate for non-METER data loggers to connect TEROS 31 and TEROS 32

ADVANTAGES

  • Digitálny senzor komunikuje viacero meraní cez sériové rozhranie
  • Nízkopríkonový objtage požiadavky
  • Dizajn s nízkou spotrebou energie podporuje batériové záznamníky údajov
  • SDI-12, Modbus RTU or tensio LINK serial communications protocol supported
  • Analog output supported (only 8-pin version)

ŠPECIFIKÁCIA

ŠPECIFIKÁCIE MERANIA
Barometrický tlak
Rozsah + 65 kPa až + 105 kPa
Rozlíšenie ± 0.0012 kPa
Presnosť ± 0.05 kPa
Teplota
Rozsah -30 až + 60 °C
Rozlíšenie ± 0.01 °C
Presnosť ± 0.5 °C
 KOMUNIKAČNÉ ŠPECIFIKÁCIE
Výstup
Analog Output (8-pin connector only)0 to 2,000 mV (default)0 to 1,000 mV (configurable with tensio VIEW)
Digital OutputSDI-12 communications protocol tensio LINK communication protocol Modbus RTU communication protocol
Kompatibilita zapisovača údajov
Analógový výstup Ľubovoľný systém na zber údajov schopný prepínaného budenia 3.6 až 28 V DC a jednostranného alebo diferenciálneho napájaniatagmeranie s rozlíšením väčším alebo rovným 12 bitom.
Digitálny výstup Ľubovoľný systém na zber údajov schopný budenia 3.6 až 28 V DC a komunikácie RS-485 Modbus alebo SDI-12.
 FYZIKÁLNE ŠPECIFIKÁCIE
Rozmery
Dĺžka 80 mm (3.15 palca)
šírka 29 mm (1.14 palca)
Výška 30 mm (1.18 palca)
Dĺžka kábla
1.5 m (standard)NOTE: Contact Customer Support if a nonstandard cable length is needed.
Typy konektorov
4-pin and 8-pin M12 plug connector or stripped and tinned wires
 SÚLAD
EM ISO/IEC 17050:2010 (značka CE)

EKVIVALENTNÉ TYPY OBVODU A PRIPOJENIA
Refer to Figure 2 to connect the BARO Module to a data logger. Figure 2 provides a low-impedance variant of the recommended SDI-12 specification.

METER-BARO-Module- (2)

METER-BARO-Module- (3)

METER-BARO-Module- (4)

PRÍRUČKA INTEGRÁTORA MODULU BARO METER-BARO-Module- (5)

METER-BARO-Module- (6)

METER-BARO-Module- (7)

METER-BARO-Module- (8)

PREVENCIA

Senzory METER sú vyrobené podľa najvyšších štandardov, ale nesprávne použitie, nesprávna ochrana alebo nesprávna inštalácia môžu senzor poškodiť a prípadne zrušiť záruku. Pred integráciou snímačov do siete snímačov postupujte podľa odporúčaných inštalačných pokynov a implementujte bezpečnostné opatrenia na ochranu snímača pred škodlivým rušením.

KOMUNIKÁCIA SENZOROM
METER digital sensors feature a serial interface with shared receive and transmit signals for communicating sensor measurements on the data wire. The sensor supports SDI-12, tensio LINK, and Modbus over RS-485 two-wire. The sensor automatically detects the interface and protocol which is being used. Each protocol has implementation advantaga výzvy. Ak nie je výber protokolu pre požadovanú aplikáciu zrejmý, kontaktujte zákaznícku podporu spoločnosti METER.

  • SDI-12 ÚVOD
    SDI-12 je protokol založený na štandardoch na prepojenie snímačov s dataloggermi a zariadeniami na zber údajov. Viaceré senzory s jedinečnými adresami môžu zdieľať spoločnú 3-vodičovú zbernicu (napájanie, uzemnenie a dáta). Obojsmerná komunikácia medzi snímačom a záznamníkom je možná zdieľaním dátovej linky pre vysielanie a prijímanie, ako je definované štandardom. Merania snímača sa spúšťajú príkazom protokolu. Protokol SDI-12 vyžaduje jedinečnú alfanumerickú adresu snímača pre každý snímač na zbernici, aby záznamník údajov mohol odosielať príkazy a prijímať údaje zo špecifických snímačov.
    Stiahnite si špecifikáciu SDI-12 v1.3 a dozviete sa viac o protokole SDI-12.
  • ÚVOD DO RS-485
    RS-485 je robustné fyzické zbernicové pripojenie na pripojenie viacerých zariadení k jednej zbernici. Je schopné používať veľmi dlhé káblové vzdialenosti v náročných podmienkach. Namiesto SDI-12 používa RS-485 dva vyhradené vodiče pre dátový signál. To umožňuje použitie dlhších káblov a je odolnejšie voči rušeniu z vonkajších zdrojov, pretože signál je prepojený s rôznymi vodičmi a napájacie prúdy neovplyvňujú dátový signál. Viac informácií o RS-485 nájdete na Wikipédii.
  • ÚVOD TENSIOLINK RS-485
    tensioLINK je rýchly, spoľahlivý, proprietárny sériový komunikačný protokol, ktorý komunikuje cez rozhranie RS-485. Tento protokol sa používa na čítanie údajov a konfiguráciu funkcií zariadenia. METER poskytuje prevodník tensioLINK PC USB a softvér na priamu komunikáciu so senzorom, čítanie údajov a aktualizáciu firmvéru. Pre viac informácií o tensioLINK kontaktujte zákaznícku podporu.
  • ÚVOD DO MODBUS RTU RS-485
    Modbus RTU je bežný sériový komunikačný protokol používaný programovateľnými logickými automatmi (PLC) alebo záznamníkmi údajov na komunikáciu so všetkými druhmi digitálnych zariadení. Komunikácia prebieha cez fyzické pripojenie RS-485. Kombinácia RS-485 pre fyzické pripojenie a Modbus ako sériového komunikačného protokolu umožňuje rýchly a spoľahlivý prenos údajov pre veľký počet senzorov pripojených k jednému vodiču sériovej zbernice. Ďalšie informácie o Modbus nájdete na nasledujúcich odkazoch: Wikipédia a modbus.org.
  • PREPOJENIE SNÍMAČA S POČÍTAČOM
    The serial signals and protocols supported by the sensor require some type of interface hardware to be compatible with the serial port found on most computers (or USB-to-serial adapters). There are several
    SDI-12 interface adapters available in the marketplace; however, METER has not tested any of these interfaces and cannot make a recommendation as to which adapters work with METER sensors. METER data loggers and the ZSC handheld device can operate as a computer-to-sensor interface for making on-demand sensor measurements.
    Modul BARO je možné konfigurovať a merať aj cez tensioLINK pomocou softvéru METER.VIEW, k dispozícii na stiahnutie na meter.ly/software. Na pripojenie modulu BARO k počítaču je potrebný USB prevodník tensioLINK a vhodný adaptérový kábel.
  • REALIZÁCIA METRA SDI-12
    Ak je medzi tenziometrom TEROS 31 alebo 32 pripojený modul BARO, je možné cez Modbus odčítať barometrický tlak vzduchu aj absolútny tlak tenziometra TEROS. Cez Modbus je možné odčítať aj kompenzovaný maticový potenciál.
    METER sensors use a low-impedance variant of the SDI-12 standard sensor circuit (Figure 2). During the power-up time, sensors output some sensor diagnostic information and should not be communicated with until the power-up time has passed. After the power up time, the sensors are fully compatible with all commands listed in the SDI-12 Specification v1.3 except for the continuous measurement commands ( aR0 – aR9 and aRC0 – aRC9 ). M , R , and C command implementations are found on pages 8–9. Out of the factory, all METER sensors start with SDI-12 address 0 .
  • ÚVAHY SENZOR BUS
    SDI-12 sensor buses require regular checking, sensor upkeep, and sensor troubleshooting. If one sensor goes down, that may take down the whole bus even if the remaining sensors are functioning normally. Power cycling the SDI-12 bus when a sensor is failing is acceptable. METER SDI-12 sensors can be power-cycled and read on the desired measurement interval or powered continuously and commands sent when a measurement is desired based on specified communication timing. Many factors influence the effectiveness of the bus configuration. Visit metergroup.com pre články a virtuálne semináre obsahujúce viac informácií.

KONFIGURÁCIA SDI-12

Tabuľka 1 uvádza konfiguráciu komunikácie SDI-12.

Tabuľka 1      Konfigurácia komunikácie SDI-12
prenosová rýchlosť 1,200
Štart Bits 1
Dátové bity 7 (LSB prvý)
Paritné bity 1 (párne)
Stop bity 1
Logika Obrátené (aktívne nízke)

ROZČASOVANIE SDI-12
All SDI-12 commands and responses must adhere to the format in Figure 9 on the data line. Both the command and response are preceded by an address and terminated by a carriage return and line feed combination ( <CR><LF> ) and follow the timing shown in Figure 10.

METER-BARO-Module- (9)

METER-BARO-Module- (10)

SPOLOČNÉ PRÍKAZY SDI-12
Táto časť obsahuje tabuľky bežných príkazov SDI-12, ktoré sa často používajú v systéme SDI-12 a zodpovedajúce odozvy zo snímačov METER.

IDENTIFIKAČNÝ PRÍKAZ (aI!)
Príkaz Identifikácia možno použiť na získanie rôznych podrobných informácií o pripojenom senzore. Bývalýampsúbor príkazu a odpovede je zobrazený v príkladeample 1, kde je príkaz tučným písmom a odpoveď nasleduje za príkazom.

Example 1 1I!113METER␣ ␣ ␣BARO␣

Parameter

Pevný znak  Dĺžka  Popis
 
1ja! 3 Data logger command.Request to the sensor for information from sensor address 1 .
1 1 Sensor address.Prepended on all responses, this indicates which sensor on the bus is returning the following information.
13 2 Označuje, že cieľový snímač podporuje špecifikáciu SDI-12 v1.3.
METER ␣ ␣ ␣ 8 Vendor identification string.( METER and three spaces  ␣ ␣ ␣ for all METER sensors)
BARO␣ 6 Sensor model string.This string is specific to the sensor type. For the BARO, the string is BARO .
100 3 Sensor version.This number divided by 100 is the METER sensor version (e.g., 100 is version 1.00).
BARO-00001 ≤13,variable Sériové číslo snímača. Toto je pole s premenlivou dĺžkou. Pri starších snímačoch môže byť vynechaný.

PRÍKAZ ZMENIŤ ADRESU (aAB!)
The Change Address command is used to change the sensor address to a new address. All other commands support the wildcard character as the target sensor address except for this command. All METER sensors have a default address of 0 (zero) out of the factory. Supported addresses are alphanumeric (i.e., A – Z , and 0 – 9 ). An exampVýstup zo snímača METER je znázornený na príkladeample 2, kde je príkaz tučným písmom a odpoveď nasleduje za príkazom.

Example 2 1A0!0

 Parameter

Pevný znak  Dĺžka        Popis
1A0! 4 Data logger command. Request to the sensor to change its address from 1 to a new address of 0 .
0 1 Nová adresa senzora. Pre všetky nasledujúce príkazy bude cieľový senzor používať túto novú adresu.

IMPLEMENTÁCIA PRÍKAZOV
V nasledujúcich tabuľkách sú uvedené príslušné príkazy Meranie ( M ), Priebežné ( R ) a Súbežné ( C ) a následné príkazy Dáta ( D ), ak je to potrebné.

IMPLEMENTÁCIA MERACÍCH PRÍKAZOV
Measurement ( M ) commands are sent to a single sensor on the SDI-12 bus and require that subsequent Data ( D ) commands are sent to that sensor to retrieve the sensor output data before initiating communication with another sensor on the bus. Please refer to Table 2 and for an explanation of the command sequence and to Table 5 for an explanation of response parameters.

Stôl 2:XNUMX! postupnosť príkazov

Príkaz odpoveď
Tento príkaz hlási priemerné, akumulované alebo maximálne hodnoty.
am! atttn
aD0! a± ± +
Komentáre When a slave TEROS tensiometer is connected, <Press> hold the barometric compensated tensiometer output. If the BARO module is used in standalone <Press> returns the current barometric pressure.
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement command is processed by the sensor, a service request a <CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until   seconds have passed or wait until the service request is received before sending the data commands. See the SDI-12 Specifications v1.3

POZNÁMKA: Meracie a zodpovedajúce dátové príkazy sú určené na použitie bezprostredne po sebe. Po spracovaní meracieho príkazu senzorom sa odošle servisná požiadavka. sa odošle zo senzora signalizujúc, že ​​meranie je pripravené. Pred odoslaním dátových príkazov buď počkajte, kým neuplynie ttt sekúnd, alebo počkajte, kým sa neprijme požiadavka na službu. Ďalšie informácie nájdete v dokumente SDI-12 Specifications v1.3.

IMPLEMENTÁCIA PRÍKAZOV SÚČASNÉHO MERANIA
Príkazy pre súbežné meranie (C) sa zvyčajne používajú so senzormi pripojenými k zbernici. Príkazy C pre tento senzor sa líšia od štandardnej implementácie príkazu C. Najprv odošlite príkaz C, počkajte stanovený čas uvedený v odpovedi na príkaz C a potom použite príkazy D na prečítanie jeho odpovede pred komunikáciou s iným senzorom.

Please refer to Table 3 for an explanation of the command sequence and to Table 5 for an explanation of response parameters.

Table 3      aC! measurement command sequence
Príkaz odpoveď
Tento príkaz hlási okamžité hodnoty.
aC! atttnn
aD0! a± ± +
NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro- cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec- onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.

NOTE: The measurement and corresponding data commands are intended to be used back to back. After a measurement commanc is pro-cessed by the sensor, a service request a<CR><LF> is sent from the sensor signaling the measurement is ready. Either wait until ttt sec-onds have passed or wait until the service request is received before sending the data commands. Please see the SDI-12 Specifications v1.3 document for more information.

IMPLEMENTÁCIA PRÍKAZOV KONTINUÁLNEHO MERANIA
Continuous Measurement ( R ) commands trigger a sensor measurement and return the data automatically after the readings are completed without needing to send a D command. aR0! returns more characters in its response than the 75-character limitation called out in the SDI-12 Specification v1.3. It is recommended to use a buffer that can store at least 116 characters.
Please refer to Table 4 for an explanation of the command sequence and see Table 5 for an explanation of response parameters.

Table 4      aR0! measurement command sequence
Príkaz odpoveď
Tento príkaz hlási priemerné, akumulované alebo maximálne hodnoty.
aR0! a± ± +
POZNÁMKA: Tento príkaz nedodržiava časovanie odozvy SDI-12. Viac informácií nájdete v časti Implementácia METER SDI-12.

NOTE: This command does not adhere to the SDI-12 response timing. See METER SDI-12 Implementation for more information.

PARAMETRE
Table 5 lists the parameters, unit measurement, and a description of the parameters returned in command responses for the BARO Module.

Tabuľka 5      Popisy parametrov
Parameter Jednotka Popis
± Kladné alebo záporné znamienko označujúce znamienko ďalšej hodnoty
a Adresa SDI-12
n Počet meraní (pevná šírka 1)
nn Počet meraní s úvodnou nulou, ak je to potrebné (pevná šírka 2)
ttt s Maximálny čas bude trvať meranie (pevná šírka 3)
Znak tabulátora
Charakter návratu vozíka
Riadkový znak
ASCII character denoting the sensor type For BARO Module, the character is ;
sériový kontrolný súčet METER
METER 6-bitový CRC

SÉRIOVÁ IMPLEMENTÁCIA MERAČA MODBUS RTU
Modbus over Serial Line is specified in two versions – ASCII and RTU. BARO Modules communicate using RTU mode exclusively. The following explanation is always related to RTU. Table 6 lists the Modbus RTU communication and configuration.

Tabuľka 6      Modbus communication characters
Prenosová rýchlosť (b / s) 9,600 XNUMX bps
Štart Bits 1
Dátové bity 8 (LSB prvý)
Paritné bity 0 (žiadne)
Stop bity 1
Logika Štandardná (aktívna vysoká)

METER-BARO-Module- (10)Figure 11 shows a message in RTU format. The size of the data determines the length of the message. The format of each byte in the message has 10 bits, including Start and Stop Bit. Each byte is sent from left to right: Least Significant Bit (LSB) to Most Significant Bit (MBS). If no parity is implemented, an additional stop bit is transmitted to fill out the character frame to a full 11-bit asynchronous character.

The Modbus application layer implements a set of standard Function codes divided into three categories: Public, User-defined, and Reserved. Well-defined public function codes for BARO Modules are documented in the Modbus Organization, Inc. (modbus.org) community.

For a reliable interaction between the BARO Module and a Modbus Master, a minimum 50ms delay is required between every Modbus command sent on the RS-485 bus. An additional timeout is needed for every Modbus query; this timeout is device-specific and depends on the quantity of the polled registers. Generally, 100ms will work fine for most of the BARO Module.

PODPOROVANÉ FUNKCIE MODBUS

Table 7 Function Definitions

Funkcia kód Akcia Popis
01 Čítanie stavu cievky/portu Číta stav zapnutia/vypnutia diskrétnych výstupov v ModBusSlave
02 Prečítajte si stav vstupu Číta stav zapnutia/vypnutia diskrétnych vstupov v ModBusSlave
03 Čítanie uchovávacích registrov Číta binárny obsah uchovávacích registrov v ModBusSlave
04 Čítanie vstupných registrov Číta binárny obsah vstupného registra (registrov) v ModBusSlave
05 Vynútená jedna cievka/port Núti jednu cievku/port v ModBusSlave zapnúť alebo vypnúť
06 Napíšte jeden register Zapíše hodnotu do uchovávacieho registra v ModBusSlave
15 Force multiple coils/ports Forces multiple coils/ports in the ModBusSlave to either on or off
16 Napíšte viacero registrov Writes values into a series of holding registers in the ModBusSlave

REPREZENTÁCIA DÁT A TABUĽKY REGISTROV
Data values (setpoint values, parameters, sensor-specific measurement values, etc.) sent to and from the BARO Module use 16-bit and 32-bit holding (or input) registers with a 4-digit address notation. The address spaces are virtually distributed in different blocks for each data type. This is an approach to the Modbus Enron implementation. Table 8 shows the four main tables used by the BARO Module with their respective access rights. Table 9 describes the sub-blocks for each different data type representation.

Upozorňujeme, že niektoré dátové záznamníky Modbus používajú adresovanie s ofsetom +1. To niekedy spôsobuje zmätok a je to založené na neplatnosti špecifikácie Modbus. Ak sa vyskytnú problémy s implementáciou vášho programu Modbus v dátovom záznamníku, vždy skúste otestovať rôzne ofsety registrov a dátové typy. Dobrou praxou na začatie testovania je použitie známej hodnoty, ako je teplota, kde je známe, akú hodnotu možno očakávať.

Table 8      Modbus Primary Tables
Registračné číslo Typ tabuľky Prístup Popis
1xxx Diskrétne výstupné cievky Čítanie/Písanie on/off status or setup flags for the sensor
2xxx Diskrétne vstupné kontakty Prečítajte si príznaky stavu senzorov
3xxx Analógové vstupné registre Prečítajte si numerické vstupné premenné zo senzora (skutočné merania senzora)
4xxx Analógové výstupné uchovávacie registre Čítanie/Písanie číselné výstupné premenné pre senzor (parametre, požadované hodnoty, kalibrácie atď.)

NapríkladampNapríklad register 3001 je prvý analógový vstupný register (prvá dátová adresa pre vstupné registre). Číselná hodnota uložená tu by bola 16-bitová premenná typu bez znamienka, ktorá predstavuje prvý meraný parameter snímača (hodnota tlaku). Rovnaký meraný parameter (hodnota tlaku) by sa dal načítať v registri 3201, ale tentoraz ako 32-bitová hodnota s pohyblivou desatinnou čiarkou vo formáte Big-Endian. Ak Modbus Master (datalogger alebo PLC) podporuje iba 32-bitové hodnoty s pohyblivou desatinnou čiarkou vo formáte Little-Endian, potom by sa ten istý meraný parameter (rovnaká hodnota tlaku) dal načítať v registri 3301. Virtuálne podbloky majú zjednodušiť úsilie používateľa pri programovaní Modbus dotazov snímačov.

Tabuľka 9      Modbus Virtual Sub-Blocks
Registračné číslo Prístup Veľkosť Sub-Table Údaje Typ
X001-X099 Čítanie/Písanie 16 bit celé číslo so znamienkom
X101-X199 Čítanie/Písanie 16 bit celé číslo bez znamienka
X201-X299 Čítanie/Písanie 32 bit float Big-Endian format
X301-X399 Čítanie/Písanie 32 bit float Little-Endian format

MAPOVANIE REGISTROV

Tabuľka 10      Vedenie registrov
41000 (41001*) Adresa zbernice Modbus
Podrobný popis Read or update the sensor’s modbus address
Typ údajov Celé číslo bez znamienka
Povolený rozsah 1 – 247
Jednotka
Komentáre Aktualizovaná adresa podriadeného zariadenia bude uložená v energeticky nezávislej pamäti senzora.

Tabuľka 11      BARO Module Input Registers
32000 (32001*) Soil Water Potential
Podrobný popis Compensated tension value from tensiometer
Typ údajov 32 bit floating Big-Endian
Povolený rozsah -200 až +200
Jednotka kPa
Komentáre Tensiometer needs to be connected as slave
32001 (32002*) Teplota pôdy
Podrobný popis High accuracy on board temperature measurement
Typ údajov 32 bit floating Big-Endian
Povolený rozsah -30 až +60
Jednotka ° C
Komentáre Tensiometer needs to be connected as slave
32002 (32003*) Napájanie snímača objtage
Podrobný popis On board supply voltage meranie
Typ údajov 32 bit floating Big-Endian
Povolený rozsah -10 až +60
Jednotka voltov
Komentáre
32003 (32004*) BARO Stav
Podrobný popis Binary status
Typ údajov 32 bit floating Big-Endian
Povolený rozsah 0/1
Jednotka
Komentáre
32004 (32005*) BARO Reference Pressure
Podrobný popis On board high accuracy barometric pressure measurement
Typ údajov 32 bit floating Big-Endian
Povolený rozsah +70 až +120
Jednotka kPa
Komentáre
Tabuľka 11 Vstupné registre barometrického modulu (pokračovanie)
32005 (32006*) tenzometra Tlak
Podrobný popis Absolute pressure value from tensiometer
Typ údajov 32 bit floating Big-Endian
Povolený rozsah -200 až +200
Jednotka kPa
Komentáre Tensiometer needs to be connected as slave
32006 (32007*) BARO Teplota
Podrobný popis On board temperature measurement
Typ údajov 32 bit floating Big-Endian
Povolený rozsah -30 až +60
Jednotka ° C
Komentáre

*Niektoré zariadenia hlásia adresy registrov Modbus s ofsetom +1. Toto platí pre CampZáznamníky bell Scientific a Dataker. Na načítanie požadovaného registra použite číslo v zátvorkách.

EXAMPPOUŽITIE DATALOGGERA CR6 A MODBUS RTU
Campbell Scientific, Inc. CR6 Measurement and Control Datalogger supports Modbus master and Modbus slave communication to integrate Modbus SCADA networks. The Modbus communications protocol facilitates the exchange of information and data between a computer/HMI software, instruments (RTUs), and Modbus-compatible sensors. The CR6 datalogger communicates exclusively in RTU mode. In a Modbus network, each slave device has a unique address. Therefore, sensor devices must be configured correctly before connecting to a Modbus Network. Addresses range from 1 to 247. Address 0 is reserved for universal broadcasts.

PROGRAMOVANIE DÁTALOVÉHO ZARIADENIA CR6
Programy bežiace na loggeroch CR6 (a CR1000) sú napísané v jazyku CRBasic, ktorý vyvinul programátor C...ampbell Scientific. Je to jazyk na vysokej úrovni navrhnutý tak, aby poskytoval jednoduchý, ale mimoriadne flexibilný a výkonný spôsob, ako inštruovať záznamník údajov, ako a kedy vykonávať merania, spracovávať údaje a komunikovať. Programy je možné vytvárať pomocou softvéru ShortCut alebo upravovať pomocou editora CRBasic, ktoré sú k dispozícii na stiahnutie ako samostatné aplikácie na oficiálnej platforme C.ampbell Scientific webstránka (www.campbellsci.com). ShortCut Software (https://www.campbellsci.com/shortcut) CRBasic Editor (https://www.campbellsci.com/crbasiceditor)

A typical CRBasic program for a Modbus application consists of the following:

  • Variables and constants declarations (public or private)
  • Units declarations
  • Konfiguračné parametre
  • Data tables declarations
  • Logger Initializations
  • Scan (Main Loop) with all the sensors to be quired
  • Function call to the Data Tables

ROZHRANIE PRIPOJENIA RS-485 ZAznamenávača CR6
Univerzálny (U) terminál CR6 ponúka 12 kanálov, ktoré sa dajú pripojiť takmer ku každému typu senzora. Umožňuje CR6 prispôsobiť sa väčšiemu počtu aplikácií a eliminuje použitie mnohých externých periférií.
Pripojenie Modbus CR6 znázornené na obrázku 12 používa rozhranie RS-485 (A/B) namontované na svorkách (C1-C2) a (C3-C4). Tieto rozhrania môžu pracovať v polovičnom duplexe a plnom duplexe. Sériové rozhranie modulu BARO použité pre tento účel...ampje pripojený ku svorkám (C1-C2).

Schéma zapojenia modulu BARO k dataloggeru CR6METER-BARO-Module- (12)

After assigning the BARO Module a unique Modbus Slave Address, it can be wired to the CR6 logger according to Figure 12. Make sure to connect the white and black wires according to their signals, respectively, to the C1 and C2 ports—the brown wire to 12V (V+) and the blue to G (GND). To control the power supply through your program, connect the brown wire directly to one of the SW12 terminals (switched 12V outputs).

EXAMPPROGRAMY LE

METER-BARO-Module- (13) METER-BARO-Module- (14) METER-BARO-Module- (15)

ZÁKAZNÍCKA PODPORA

SEVERNÁ AMERIKA
Zástupcovia zákazníckeho servisu sú k dispozícii v prípade otázok, problémov alebo spätnej väzby od pondelka do piatku od 7:00 do 5:00 tichomorského času.

EURÓPA

Ak kontaktujete METER e-mailom, uveďte nasledujúce informácie:

  • Meno
  • Adresa
  • Telefónne číslo
  • E-mailová adresa
  • Sériové číslo prístroja

Popis problému

POZNÁMKA: V prípade produktov zakúpených prostredníctvom distribútora sa so žiadosťou o pomoc obráťte priamo na distribútora.

HISTÓRIA REVÍZIE
Nasledujúca tabuľka obsahuje zoznam revízií dokumentov.

Revízia Dátum Kompatibilný firmvér Popis
00 6.2025 1.10 Prvotné uvoľnenie

FAQ

What should I do if I need a nonstandard cable length?

Contact Customer Support for assistance with nonstandard cable lengths.

How do I know which communication protocol to use for my application?

Evaluate the advantages and challenges of each protocol based on your application needs. If unsure, contact METER Customer Support for guidance.

Dokumenty / zdroje

PDF thumbnailModul BARO
User Guide · TEROS 31, TEROS 32, BARO Module, BARO Module, Module

Položiť otázku

Use this section to ask about setup, compatibility, troubleshooting, or anything missing from this manual.

Položiť otázku

Ask about setup, compatibility, troubleshooting, or anything missing from this manual. Name and email are optional.