TSI-LOGO

TSI OmniTrak Solution Core Module

TSI-OmniTrak-Solution-Core-Module-PRODUCT

User Calibration Guide

  1. Safety & Handling Notes
    • Use calibration gases in a well-ventilated area.
    • Do not exceed recommended gas concentrations or pressures.
    • Secure gas cylinders upright and use appropriate regulators.
    • Always wear PPE when handling compressed gases.
  2. Kedy kalibrovať
    • Factory calibration – ensures initial accuracy
      • Ročne
    • Field calibration – compensates for real-world influences and aging over time
      • At least daily or before each study
      • More frequently if:
        • Readings appear inaccurate or drift beyond expected range.
        • It’s exposed to high VOC or CO2 environments, high humidity, dust, or pollutants.
        • Regulatory or certification programs require it.
  3. Požadované vybavenie
    • OmniTrak™ Core module calibration shroud (7590-84)TSI-OmniTrak-Solution-Core-Module-FIG-1
    • Zero air source and/or span gas (see Recommended Gases and Concentrations table)
    • Gas regulator for ≤ 0.3 L/min
    • FEP lined tubing and regulator
    • OmniTrak™ Smart Station
  4. Pokyny na nastavenieTSI-OmniTrak-Solution-Core-Module-FIG-2
    • Start with the Core module paired to the Smart Station
    • Attach the calibration shroud to the Core Module next to the probe
    • Connect tubing from gas source to the calibration shroud
    • On the Smart Station, tap the Connection menu (network icon in the top right corner). Find the Core module to calibrate and tap the kebab (three vertical dots) menu next to the module. Select ‘Calibrate’ and follow the instructions.
  5. Kalibrácia
    For the VOC-LOW sensor, Zero Calibration and Span Calibration are available. If both are performed, Zero Calibration should be completed before Span Calibration.TSI-OmniTrak-Solution-Core-Module-FIG-3

VOC Zero Calibration

  • On the Smart Station Core Module Calibration screen:
  • Select ‘VOC LOW’
  • Tap ‘Calibrate’
  • Tap ‘Zero’
  • Apply the zero gas bottle. Tap ‘Continue.’
  • Flow zero air into the sensor at ≤ 0.3 L/min
  • Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
  • Wait for the confirmation message that the zero calibration was successful.

TSI-OmniTrak-Solution-Core-Module-FIG-4

VOC Span Calibration

  • On the Smart Station Core Module Calibration screen:
  • Select ‘VOC LOW’
  • Tap ‘Calibrate’
  • Tap ‘Span’
  • Flow VOC span gas (i.e., 10,000 ppb isobutylene) into the sensor at ≤ 0.3 L/min
  • Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
  • Enter Span Gas Concentration (ppb)
  • Wait for confirmation message that the span calibration was successful.

TSI-OmniTrak-Solution-Core-Module-FIG-5

CO2 Span Calibration

  • On the Smart Station Core Module Calibration screen:
  • Select ‘CO2’
  • Tap ‘Calibrate’
  • Tap ‘Span’
  • Flow CO2 span gas (i.e., 400 ppb CO2) into the sensor at ≤ 0.3 L/min
  • Wait at least 2 minutes for reading to stabilize, then press ‘Apply’
  • Enter Span Gas Concentration (ppm)
  • Wait for confirmation message that the span calibration was successful.

Post-Calibration Verification

  • Flow gas again and confirm sensor is reading within +/- 10% of target
  • Return to ambient air and confirm baseline VOC and CO2 return to expected values.

Riešenie problémov

Vydanie Možná príčina Riešenie
Kalibrácia zlyhala Nestabilné hodnoty Let readings stabilize before retrying
VOC reading too high post-zero Contamination in zero air source Replace or verify zero air source
CO2 span fails Incorrect gas concentration Verify gas cylinder and regulator

Dodatok

Recommended Gases and Concentrations

Typ snímača Typ kalibrácie Gas Used Účel
 

 

VOC

nula Air with zero VOCs/reference zero air, ≤ 0.3 L/min Establish baseline free of ambient VOCs
 

Span

The known VOC of interest, isobutylene in air is recommended. <10,000 ppb (10 ppm) recommended, ≤ 0.3 L/min Ensure correct sensitivity to target VOCs
CO2 Span CO2 in air, 400 to 2000 ppm recommended,

≤ 0.3 l/min

Ensure accurate readings at expected levels

TSI Technical Support

Please contact TSI or your local channel partner to make service arrangements and to receive a Service Request number. To fill out an online Service Request form, visit TSI Incorporated’s webmiesto na tsi.com/support.

TSI začlenená

500 Cardigan Road
Pobrežieview, MN 55126

Tel:……….+1 800-680-1220
……….+1 651-490-2860
Fax:………+1 651-490-3824

TSI Incorporated – Navštívte nás webstránky www.tsi.com pre viac informácií.

TSI-OmniTrak-Solution-Core-Module-FIG-7

TSI and the TSI logo are registered trademarks of TSI Incorporated in the United States and may be protected under other countries’ trademark registrations.

  • USA Tel: +1 800 680 1220
  • UK Tel: +44 149 4 459200
  • Francúzsko Tel: +33 1 41 19 21 99
  • Nemecko Tel: +49 241 523030
  • India Tel: +91 80 67877200
  • Čína Tel: +86 10 8219 7688
  • Singapur Tel: +65 6595 6388

P/N 6018700 Rev. A US (10/7/2025)

©2025 TSI Incorporated Vytlačené v USA

Dokumenty / zdroje

TSI OmniTrak Solution Core Module [pdfPokyny
30771, 7590-84, OmniTrak, Module, OmniTrak Solution Core Module, Solution Core Module, Module

Referencie

Zanechajte komentár

Vaša emailová adresa nebude zverejnená. Povinné polia sú označené *